Thank you for purchasing this OMRON Professional Blood Pressure Monitor.
Please completely read this Instruction Manual before using the monitor for the first time.
Read this manual to ensure the safe and accurate use of the monitor.
AR
Przeglądanie stron 0
1 2 3 4 5 6 ... 37 38

Podsumowanie treści

Strona 1 - HBP-1300

Professional Blood Pressure MonitorHBP-1300• Instruction Manual• Mode d’emploi• Gebrauchsanweisung• Manuale di istruzioni• Manual de instrucciones• G

Strona 2 - Contents

8Non-Invasive Blood Pressure (NIBP) measurement Warning CautionIf a cuff is used on a patient with an infection, treat the cuff as medical waste, or d

Strona 3 - Introduction

9ENNote:Setup• Read and understand the manual for each optional accessory. This manual does not contain cautionary information for optional accessory.

Strona 4 - General Information

10Using the UnitComponents of the ProductBefore using the unit, make sure that no accessories are missing and that the unit and accessories are not da

Strona 5

11ENOptional Medical Accessories(within the scope of EC Medical Device Directive 93/42/EEC)* UK plug type AC ADAPTER-UK1600(9994843-9) CautionOnly OMR

Strona 6

12Features / Functions of UnitFront and back of unitMeaning of the Symbols1[START/STOP] button(Power ON/OFF)Power ON / Measure blood pressurePress whe

Strona 7

13ENLCD Display* Only when the battery is installed.1 SYS Displays systolic blood pressure.2 DIA Displays diastolic blood pressure.3 Pulse Displays th

Strona 8

14Installing the Battery Pack CautionDo not short the positive and negative terminals of the battery pack with a wire or other metal object. If batter

Strona 9

15ENBattery chargingBattery levelConnecting the AC AdapterAC powerVerify that the power outlet supplies the specified voltage and frequency (100 - 240

Strona 10 - Warning

16System SettingsThe system settings are divided into two modes, “Menu Mode” and “Utility Mode”.Menu Mode“Menu Mode” allows you to configure the “Init

Strona 11 - Rechargeable battery

17ENUtility Mode“Utility Mode” allows you to configure the “Auto Power Off” and “Pressure Accuracy Confirmation” setting.1. Confirm that the device is

Strona 12 - Using the Unit

ContentsIntroductionIntended Use ...1Exe

Strona 13 - Caution

18Non-Invasive Blood Pressure (NIBP) MeasurementNon-Invasive Pressure Measurement PrinciplesOscillometric methodThe beat in the pulsation generated by

Strona 14 - Features / Functions of Unit

19ENCuff Selection and ConnectionSelecting the cuff Caution• Do not use the cuff if it is damaged or has holes.• Use the appropriate cuff size to ensu

Strona 15 - LCD Display

20Applying the Cuff to the Patient1. Wrap on a bare arm or over thin clothing.Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a r

Strona 16

21EN CautionMake sure the cuff is wrapped in the correct arm position and is at the same height as the heart. A difference of 10 cm (4 inches) in heig

Strona 17 - Connecting the AC Adapter

22Irregular pulse wave detection functionIf the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection icon will

Strona 18 - System Settings

23ENStopping the MeasurementTo stop measurement while measurement is in progress, press [START/STOP] button.Displaying Last ReadingThe previous measur

Strona 19

24MaintenanceMaintenance Inspection and Safety ManagementThe HBP-1300 must be maintained to ensure functionality and to secure the safety of patients

Strona 20 - OSCILLATIONS IN CUFF PRESSURE

25ENCheck before UseBefore conducting safety checks, be sure to implement the items in the “Cleaning of the Device” and “Accessory Care” sections. (pa

Strona 21

26TroubleshootingThe power does not turn onCause SolutionIf the unit is being powered by the battery, the battery is not installed or the charge is de

Strona 22 - 1 to 2 cm

27ENAbnormal measurement valueCause / solutionThe causes below are possible. Check the patient by palpation and then repeat measurement.• Body movemen

Strona 23

1ENIntroductionIntended UseMedical PurposeThe device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult and ped

Strona 24

28List of Error CodesThe alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority alarm occurs.To clear an alar

Strona 25 - Displaying Last Reading

29ENOther problemsDisposalDescriptionAs there is a risk of environmental pollution, follow your applicable national and local legal regulations regard

Strona 26 - Maintenance

30SpecificationsFactory Default SettingsFactory default settings and backup are as shown below.Backup: Setting is retained even if the power is interr

Strona 27 - Check before Use

31ENEnvironmental ConditionsNon-Invasive Blood Pressure (NIBP)* Comparison with auscultation method performed by a trained professional.DIA determined

Strona 28 - Troubleshooting

32Important information regarding Electro Magnetic Compatibility (EMC)With the increased number of electronic devices such as PC’s and mobile (cellula

Strona 29 - Stethoscope

33ENManufacturer’s DeclarationThe HBP-1300 is intended for use in the electromagnetic environment specified below. The customer or the user of the HBP

Strona 30 - List of Error Codes

34Immunity testIEC60601-1-2 test levelCompliance levelElectromagnetic environment - guidanceConducted RFIEC 61000-4-6Radiated RFIEC 61000-4-33 Vrms150

Strona 31 - Disposal

35ENRecommended Separation Distances:Recommended separation distance between portable and mobile RF communications equipment and the HBP-1300The HBP-1

Strona 32 - Specifications

Made in ChinaManufacturer OMRON HEALTHCARE Co., Ltd.53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPANEU-representative OMRON HEALTHCARE EUROPE B.

Strona 33 - Environmental Conditions

2Notes on SafetyThe warning signs and symbol examples indicated below are intended to ensure safe use of the product and prevent damage and injury to

Strona 34

3ENWarnings and CautionsUsage warnings and cautionsSetup Warning CautionDo not use the cuff or AC adapter to lift the unit, it can also cause the unit

Strona 35 - Manufacturer’s Declaration

4Before use / during use WarningThe unit complies with the EMC standard (IEC60601-1-2). As such, it can be used simultaneously with multiple medical i

Strona 36

5EN CautionBefore using the unit, verify that none of the following apply to the patient:- Poor peripheral circulation, noticeably low blood pressure,

Strona 37

6Cleaning Warning CautionMaintenance and inspection WarningWhen cleaning the unit, turn off the power and disconnect the AC adapter from the unit.Afte

Strona 38

7ENWarnings and cautions for safe measurementRechargeable battery Warning CautionIf battery fluid comes in contact with the eye, immediately flush wit